ChemLogic® CL96 Operating Manual
Appendix F Hard Wire Connection (Optional)
WARNING :The detachable power cord or the supply line wiring must meet the
rating specified below .
When the hard-wired option is specified the installation must be done by a qualified electrician.
IMPORTANT: When installing A/C voltage to the CL96 for the hard wire connection read carefully:
A switch or circuit-breaker shall be included in the building installation.
It shall be in close proximity to the equipment and within easy reach of the OPERATOR.
It shall be marked as the disconnecting point for the CL96 equipment.
When the Hard-Wire option is specified the A/C power must be provided by the customer to the FILTER which
is located on the right side of the unit inside the back cabinet. After disconnecting all power to the CL96
remove the back cabinet which will reveal the A/C filter shown in figure 1.
NOTE THE A/C FILTER IS MOUNTED UPSIDE DOWN IN THE CL96
Connect A/C power to the screw terminals numbered 1 & 2
Screw # 1 is Line and Screw #2 is neutral
Connect the ground wire to the terminal as shown in Figure 2
Replace the protective plastic cover over the screws
Make sure the wires are tightened properly and the protective plastic covers are replaced over the screws.