7.7 Flow Adjustment
Each channel should be adjusted whenever a new ChemLogic® tape or particulate filter is installed. See
section 6.2.b.
7.8 USB Storage Drive Replacement
It is highly recommended to keep a USB flash drive inserted in the unit at all times. A general fault is issued
anytime a USB drive is not inserted or full. To insert or replace a USB drive follow this steps.
1. Exit analysis
2. Go to the USB Drive removal screen on the main menu under History.
3. Touch the ‘remove’ button to halt writing to the disk.
4. Follow the procedures from section 7.2 to open the control panel access door.
5. Remove the USB drive from the system and replace with new drive.
6. Wait 5 seconds for the system to initialize.
7. Close and latch the control panel access door.
Appendix F details the data stored on the CF card and how to access it on a personal computer.