Instructions for use Babyleo TN500 SW 1.0n 59
Assembly and preparation
Fig. 29 Correct side of device for oxygen cylinders
WARNING
Risk of explosion and fire
If oxygen cylinders or devices containing oxygen are installed in the vicinity of the
ventilation slot, oxygen may get into the device at the power supply unit. In the
event of sparking at the power supply unit there is an increased risk of fire, which
may lead to personal injury and property damage.
► Position oxygen cylinders and devices containing oxygen on that side of the
device without the ventilation slot.
WARNING
Risk to patient if unauthorized gases are used
The use of compressed gases that are not approved for medical purposes may
adversely affect the function of the device and put the patient's health at risk.
► Only use compressed gases that are approved for medical use.
► Only use compressed gases that are dry and free from dust and oil.
WARNING
Increased risk of fire from unauthorized oxygen cylinder valves
The use of unauthorized or poorly maintained oxygen cylinder valves presents an
increased risk of fire, which may lead to personal injury and property damage.
► Do not use unauthorized oxygen cylinder valves.
► Make sure that oxygen connectors and oxygen seals are kept free from oil and
dust at all times.
► Always open oxygen cylinder valves slowly.