60 Instructions for use Babyleo TN500 SW 1.0n
Assembly and preparation
WARNING
Risk of crushing if accessories are mounted at too low a height
If compressed gas cylinders are present and accessories are mounted too low
down on the duct rail, hands or fingers may be crushed between the duct rail and
the compressed gas cylinders.
► If compressed gas cylinders are present, do not mount accessories too low
down.
► When lowering the height adjustment, do not touch the compressed gas
cylinders and the accessory arm.
5.3.1 Adjusting the gas cylinder holder
Before installation, set the lifting column to its lowest position. Mount
accessories with the lifting column in its lowest position. Do not mount
accessories too low down.
Procedure:
1 Adjust the bottom of the gas cylinder holder (3) to the height of the oxygen
cylinder. The height of the gas cylinder holder can be set to 2 levels.
2 Ensure that the pressure reducer is pointing towards the rear and protrudes over
the upper edge of the gas cylinder holder.
3 Adjust the height of the protection bar (2) to the size of the oxygen cylinder and
the position of the pressure reducer.
Fig. 30 Adjusting the height of the gas cylinder holder
4 Fit the spacers to the duct rails (1) to prevent the accessories colliding with the
oxygen cylinder when the device is lowered.
✓ The gas cylinder holder has been adjusted.
30227