Instructions for use Babyleo TN500 SW 1.0n 87
Operation
6.9.2 Starting air temperature mode
This chapter describes how to start air temperature mode.
Prerequisites:
– The device is in operation.
– The hood is closed.
WARNING
Risk of hyperthermia due to external heat sources
The temperature in the patient compartment may rise to dangerous levels due to
direct sunlight or other external heat sources (e.g., phototherapy devices). The
patient may become overheated.
► When additional heat sources (e.g., solar radiation, examination lights, or
phototherapy lights) are active, make sure that the temperature in the patient
compartment remains non-critical.
► Check the patient's body core temperature at regular intervals.
WARNING
Risk to patient due to incorrect settings
If the air temperature and humidity settings are unsuitable for the patient, the patient
may be put at risk.
► Check the suggested air temperature and humidity settings carefully and adjust
them to the patient's needs if necessary.
Procedure:
1 Touch the Air mode tab (1) in the therapy bar.
The tab turns yellow.
2 Press the rotary knob to confirm.
The tab turns dark green.
The mode is activated.
Fig. 45 Starting air temperature mode