Instructions for use Babyleo TN500 SW 1.0n 93
Operation
Procedure:
1 Touch the Manual mode tab (1).
The tab turns yellow.
2 Press the rotary knob to confirm.
The tab turns dark green.
Fig. 49 Starting manual mode
3 Touch the Warmer therapy control (2).
The therapy control turns yellow.
4 Set the radiant warmer power using the rotary knob.
5 Press the rotary knob to confirm.
The therapy control turns dark green.
The radiant warmer power is set.
If the radiant warmer power is set above 30 %, a timer starts. The device
switches the radiant warmer off after 15 minutes.
6.10.3 Setting the advanced skin temperature alarm limits
Provided the following prerequisites are met, alarms for the skin temperature can
be set in manual mode:
– The skin temperature sensors are connected.
– The Advanced skin temperature monitoring function is activated.
For further information, see the following chapter: ''Setting the advanced skin
temperature alarm limits'' (page 84)
27253