14 Instructions for use | Oxylog VE300 SW 1.n
Safety-related information
2.8.13 Use of the rechargeable battery
Observe the following instructions when using the rechargeable battery:
► Remove the battery if the device is not going to be used for an extended period
of time.
► In the absence of the rechargeable battery, the integrated clock is powered by a
button cell. In the event of permanent storage without the rechargeable battery,
replace the button cell after 2 years.
► Fully charge the battery if the device has not been used for an extended period
of time.
► Always fit the battery correctly in the device, applying only a small amount of
force to do so.
► Before each use, make sure that the battery is sufficiently charged.
► Always keep a fully-charged battery in reserve.
► Do not throw the battery into a fire or try to open by force.
2.8.14 Cleaning and service
Disinfectants and cleaning agents
Residues of disinfectants or cleaning agents left in the device after reprocessing
may put the patient at risk due to irritation to the skin or mucous membranes.
► Observe the instructions relating to disinfectants and cleaning agents.
► Make sure that the device is adequately ventilated after reprocessing.
Risk of infection
The device must be reprocessed, otherwise there is an increased risk of infection
and the function of the device may be impaired.
► Observe the hygiene regulations at the operating location (e.g., hospital or
ambulance station), including the reprocessing intervals and reprocessing
procedures.
► Have reprocessing personnel perform reprocessing with validated procedures.
► Reprocess reusable products before their first use and after each use.
► Follow the manufacturer's instructions in respect of cleaning agents and
disinfectants.
► Reprocess the product before performing service activities or returning it for
repair. For further information, see the following chapter:
"Reprocessing", page 98.
Service
This product must be maintained at regular intervals and in the appropriate manner.
Failure to perform service activities correctly may result in personal injury and
property damage.
► Perform service activities. For further information, see the following chapter:
"Service", page 106.
► Service must be performed by those target groups that are assigned to the
particular measure.
► Only carry out service activities when there is no patient connected to the
device.