62 Instructions for use | Oxylog VE300 SW 1.n
Operation
1. Adjust the following parameters using the therapy controls:
A Maximum airway pressure Pmax
B Positive end-expiratory pressure PEEP
2. Press the More settings button (C) and adjust the following parameters and
settings:
A Inspiratory oxygen concentration FiO
2 (100 % O2 or O2/air mix)
B Non-invasive ventilation NIV (On or Off)
C Pressure support ∆Psupp
D Trigger sensitivity Trigger
E Pressure rise time Slope
SPN-CPAP/PS
The following parameters can be adjusted on-screen in addition to the parameters
of ventilation mode SPN-CPAP:
– Pressure support ∆Psupp above PEEP (C)
– Trigger sensitivity Trigger (D)
– Pressure rise time Slope (E)
The parameter Trigger can only be changed when a value > 0 mbar has been set
for ∆Psupp.
The parameter Slope can then be changed:
– Flat = slow pressure rise
– Medium = medium pressure rise
– Steep = fast pressure rise
7.3.5 Cardiopulmonary resuscitation (CPR)
WARNING
Risk of patient injury
In the ventilation category CPR the minute volume and respiratory rate alarms are
not available.
► Only use the ventilation category CPR for reanimation.
During cardiopulmonary resuscitation, the airway pressure Paw is increased
because of chest compressions.
The airway pressure Paw is monitored by the device and limited to the set
maximum inspiratory pressure Pmax. Pmax is not exceeded and the inspiration is
not ended prematurely.