Instructions for use | Oxylog VE300 SW 1.n 89
Troubleshooting
!!! Ambient
pressure low
The device measures
an ambient pressure
below the permissible
ambient pressure.
Move the device to an
environment with
higher air pressure.
15
Technical fault. Disconnect the device
from the patient and
continue ventilation
immediately using
another ventilator.
Contact DrägerService.
!!! Apnea The patient stopped to
breathe spontaneously
or is disconnected from
the device.
Check the patient's
condition.
12
Ventilate in VC-CMV
mode.
Ensure that the hose
connections are tight.
!!! Battery
discharged
The battery is dis-
charged and no exter-
nal power supply is
connected.
The ventilator must
immediately be con-
nected to the external
power supply or an on-
board power supply, or
a fully-charged battery
must be installed.
7
! Battery
inspection
due
The scheduled service
interval for the battery
has expired.
Replace the battery.
Contact DrägerService.
51
!! Battery low The device draws its
power from the battery
due to the absence of
an external power sup-
ply. There is only a
duration of operation of
approximately
15 minutes remaining in
the battery.
The ventilator must
immediately be con-
nected to the external
power supply or an on-
board power supply, or
a fully-charged battery
must be installed (venti-
lation stops while the
battery is installed).
39
! Battery not
detected
The battery is not
installed or faulty, or a
wrong battery is
installed.
Insert or replace the
battery.
42
! Battery tem-
perature high
The battery tempera-
ture is above the maxi-
mum operating
temperature.
Move the device to a
cooler environment.
40
! Battery tem-
perature low
The battery tempera-
ture is below the mini-
mum operating
temperature.
Move the device to a
warmer environment.
41
Alarm
prior-
ity
Alarm Cause Remedy Inter-
nal pri-
ority