A stenosis can also cause the filling time of the lungs to
become longer than the set inspiration time T
insp.
Starting behavior of AutoFlow
Upon switching on the function AutoFlow Savina applies
the set tidal volume VT using a volume-controlled breathing
stroke.
The plateau pressure Pplat derived from this breathing stroke
serves the function AutoFlow as starting value for the
inspiration pressure.
Settings to suit the patient’s needs
The Start of the mandatory inspiration can be synchronized
with the patient’s inhalation effort with the help of the variable
flow trigger. The flow trigger can only be completely turned off
in IPPV (IPPVAssist –> IPPV).
The steepness of the pressure increase from PEEP level to the
inspiration level can further be adapted with the ventilation
parameter FlowAcc in SIMV and IPPV to suit the patient’s
needs.