Description
107
Automatic volume compensation
If volume-controlled ventilation modes (IPPV, SIMV) are
applied in »Mask/NIV« mode, Savina increases the applied
tidal volume V
T by the amount lost due to leakage during
inspiration. If a tidal volume VT of 500 mL has been set and
50 mL are lost during inspiration, Savina will automatically
deliver 550 mL. Because unlimited leakage compensation is
not advisable, Savina compensates for leakage up to 100 % of
the set tidal volume.