Operation
22
Flow Acceleration FlowAcc
By setting the FlowAcc, the pressure and flow increase at the
start of inspiration can be changed.
To set FlowAcc:
Press »Settings ∆∆«key,
screen page »Settings 1/1« appears.
Select line »FlowAcc« on screen = turn rotary knob,
to confirm = press rotary knob.
To set value = turn rotary knob
Setting range: 5 to 200 mbar/sec
FlowAcc high: steep pressure and flow increase
To confirm setting = press rotary knob.
A recommended start-up setting for FlowAcc is 35 mbar/s.
IPPV can be extended using the following ventilation
parameters:
Trigger
Sigh
AutoFlow (optional)
Trigger (IPPVAssist)
To synchronize the mandatory ventilation strokes with the
patient’s efforts to breathe in. The actual frequency may be
higher than the ventilation frequency f set.
1 The green LED lights up if the Savina detects attempted
spontaneous breathing by the patient.
The trigger can be switched off if spontaneous breathing
by the patient is not expected or desired.
To set the trigger, see page 23.