Operation
24
Deactivating the sigh
● Set a value less than 1. OFF appears in the "∆ Sigh",
to confirm = press rotary knob.
The sigh function is deactivated.
AutoFlow® (optional)
Automatic optimization of the inspiration flow.
The inspiration flow is decelerated and regulated by
AutoFlow
*
, so that at the selected tidal volume VT, with the
current compliance a minimum airway pressure is reached and
pressure peaks are avoided.
Savina supplies additional inspiration flow when the patient
breathes in – restricted by the alarm limit VTi >.
The patient can also breathe out during the inspiratory plateau
phase.
The inspiratory pressure is limited by the Paw > alarm limit.
To switch AutoFlow ON and OFF
● Press »Settings ∆∆« key.
● Select line »AutoFlow« on screen = turn rotary knob.
● To release »AutoFlow« for setting = press rotary knob.
● To set »AutoFlow ON« = turn rotary knob,
to confirm = press rotary knob.
● To set alarm limits, see page 30.
* For a detailed description of AutoFlow, page 100.