Fault – Cause – Remedy
65
! Tidal volume low After second breathing stroke:
Because of pressure and time limitations
the set tidal volume VT will not be
applied.
Extend inspiration time »Tinsp«,
raise »FlowAcc«,
raise pressure limitation using »Pinsp«
key.
After second breathing stroke in
AutoFlow:
Because of pressure and time limitations
the set tidal volume VT will not be
applied.
Extend inspiration time »Tinsp«,
raise »FlowAcc«,
raise upper alarm limit for airway pressure
Paw .
! VTi high Alarm off
Only in »Mask/NIV« mode (optional)
The inspiratory tidal volume
upper alarm limit VTi has been
switched off.
Reset alarm limit VTi to desired
monitoring value if necessary.
Message Cause Remedy