Connectivity and Accessories
68 EchoNous Bladder & Vein User Guide
MODO Stand setup
For installation instructions on MODO Stand, follow the instructions shipped with the product.
To configure the system for the Samsung tablet with MSO, refer to the Stand Upgrade Kit Guide (P005465).
To configure the system for the Samsung tablet without the MSO, refer to the Modo Stand Upgrade Kit Guide
(P006018).
All the quick start guides are shipped with the product and are available in the pen drive.
EchoNous AI Station setup
For installation instructions on the AI station, refer to the AI Station Assembly Guide (P005339).
To configure the system for mobile use, refer to the AI Station Tablet Setup Guide (P005447).
Both quick start guides are shipped with the product and are available in the pen drive.
Mounting the EchoNous System display on a mobile stand is
configuring a medical electrical system. Only use an EchoNous
provided mobile stand accessory.
The ultrasound probes are connected to an Android display running
EchoNous software to configure a medical system. The display has
been certified by EchoNous as part of a medical electrical system to
EN IEC 60601-1 Edition 3.1.
Do not connect the EchoNous System display to external computers
or peripherals using the USB port unless the System is outside the
patient area. Failure to comply with these guidelines may result in
electric shock.
Only connect accessories or items that are specified as being
compatible with the EchoNous System. Contact EchoNous customer
support or your EchoNous distributor for information on compatible
accessories and systems.
Connecting electrical equipment to an MSO effectively leads to
creating a medical electrical system and can result in a reduced level
of safety.
MSOs provided with the medical electrical system (P004013) are to be
used only for supplying power to the tablet display and optional
printer in non-operating mode.
Risk of shock or personal injury when connecting any equipment that
has not been supplied as a part of the medical electrical system to the
MSO.
Do not connect an additional MSO or extension cord to the medical
electrical system.
MSOs provided with the medical electrical system (P004013) (if
provided with the medical electrical system) should only be used for
supplying power to equipment that is intended to form part of the
medical electrical system.