EasyManua.ls Logo

Ecoray HF-525Plus - Page 179

Ecoray HF-525Plus
233 pages
Print Icon
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
179
Technical Publication Document NO: SYSTEM-007-ENG (Ver.1.1)
HF-525Plus Installation, Operation and Service Manual
7.3.3 Collimator
The collimator is a device that adjusts the X-ray to expose only the part of the patient's body that is
absolutely necessary to prevent the patient from being unnecessarily exposed to the X-ray when the
X-ray is taken. The configuration of the collimator consists of an X-ray emitter, an X-axis diaphragm, a
Y-axis diaphragm, a lamp operation switch, and a DAP meter.
How to use
Figure1
(1) This is connected to the emission of the X-ray tube.
Figure2
(1) Turn this aperture knob to adjust the irradiated area of the X-ray.
(2) Adjust the Y-axis knob (Figure 2), hold the knob (Figure 2) and turn it clockwise to increase the
irradiation area. Turn it counterclockwise to reduce the irradiation area.
Figure3
(1) Before the X-ray irradiation, the lamp is operated as a method for confirming the X-ray irradiation
area, and it is the lamp operation switch (Figure 3).
(2) The operation of this collimator lamp is "ON" by pressing the operation switch (Figure 3). The lamp
operation is "ON" for 30 seconds after the switch operation and automatically "Off" after 30
seconds.
Figure4 (Optional)
(1) When the DAP meter (Figure 4) is applied, it is fixed at the end of the collimator.
Figure5
(1) Turn this aperture to adjust the irradiated area of the X-ray.
(2) Adjust the X-axis knob (Figure 5), hold the knob (Figure 5) and turn it clockwise to increase the
irradiation area and counterclockwise to reduce the irradiation area.
Collimator operation
Allow the X-ray to enter the Bucky of the Table Bucky and Wall Bucky Stand with the collimator in a
vertical position.

Table of Contents

Related product manuals