HF-525Plus Installation, Operation and Service Manual
12.4.12 Error AEC error (XR7-MiO Boar and XR7-CON Board)
1) Connect part
2) Check procedure
(1) Check Ion Chamber connector condition.
(2) Chamber position and Bucky position must be same. (Same position)
(3) Check Number flickering or not on the Time display of console after X-ray exposure with ‘beep’ sound (3
times).
(Normal: Changed normal display condition after Number flickering on the Time display)