Serinus 44  User Manual  3.0 
 
 
 
Figure 60 – Optical Filter Removal and Cleaning 
7.  Clean the filter with a lint free tissue with distilled water if further cleaning is required use high 
purity isopropyl alcohol. 
8.  Dismantle the reaction cell and clean all the stainless steel parts with high purity isopropyl 
alcohol. Then wash in an ultra-sonic bath with lab detergent and water. 
9.  Blow out all the liquid with clean oil free compressed air and dry by gently heating the assembly. 
10. Reassemble the reaction cell, replacing the 3 sintered filters (refer to Section 9.7) and test for 
leaks. 
11. Remove the black electrical tape from the optical cell while replacing the reaction cell 
immediately. 
12. Perform a full system leak test. 
13. Run a span point and calibrate the instrument. 
6.3.8  Clean Pneumatics 
The calibration valve manifold and auxiliary valve manifold will require disassembling and cleaning. 
Ideally the valves and the manifolds should be cleaned in a sonic bath with lab detergent and water. 
Once clean rinse with distilled water and dry before reassembling (refer to Section 9.9 Section 9.10 
and Section 9.11). A leak test on each assembly should be performed before adding the assembly 
back into the system. Once the system has been reassembled the instrument should have a full 
system leak check (refer to Section 6.3.5).  
If the tubing shows signs of significant contamination, it should be replaced with new tubing. 
6.3.9  Pressure Sensor Check 
Pressure checks are needed to ensure that the pressure sensor is accurately measuring pressure 
inside the instrument. 
During normal operation ensure that the Pressure & Flow Menu indicate the following parameters. 
Ambient should display the current ambient pressure at site. Cell should indicate current cell 
pressure depending on the pump condition and location. A Cell pressure of between 50 torr and 200 
torr is recommended. The manifold pressure is normally about 20 torr below ambient. 
Equipment Required 
  Barometer 
  Digital Multimeter (DMM)