EasyManua.ls Logo

Ecotech Serinus 44 - Calculation Factors Menu

Ecotech Serinus 44
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Operation
Page 55
3.4.19 Calculation Factors Menu
Main Menu Service Menu Calculation Factors Menu
The calculation factors provide the values used to calculate different aspects of measurement and
calibration.
Dilution Ratio
Entering a value here will multiply the displayed and recorded
measurements by the dilution amount. For example, if the instrument
is measuring a source where the average concentration is above the
upper limit of the measurement range, a dilution probe with a fixed
ratio dilution ratio can be used to reduce the level measured by the
instrument, so for a 4:1 dilution ratio, enter a value of 4.
Enter the ratio here so the analyser can display the correct value on the
instrument.
The default value is 1.00 (this indicates no dilution is applied).
Instrument Gain
A multiplication factor used to adjust the concentration measurement
to the appropriate level (set at calibration).
Zero Offset NO
This field displays the offset created from a NO zero calibration. This is
the concentration measured from zero air and is subtracted from all
readings.
Zero Offset NO2
This field displays the offset created from a NO
2
zero calibration. This is
the concentration measured from zero air and is subtracted from all
readings.
Zero Offset NH3
This field displays the offset created from a NH
3
zero calibration. This is
the concentration measured from zero air and is subtracted from all
readings.
Background NO
The correction factor calculated from the background cycle (used to
eliminate background interferences).
PTF Correction NO
Displays the correction factor applied to the concentration
measurement. This correction is for changes in pressure, temperature
and flows since the last calibration.
PTF Correction NOx
Displays the correction factor applied to the concentration
measurement. This correction is for changes in pressure, temperature
and flows since the last calibration.
PTF Correction Nx
Displays the correction factor applied to the concentration
measurement. This correction is for changes in pressure, temperature
and flows since the last calibration.
Conv. Efficiency NO2
The overall efficiency of the conversion of NO
2
to NO, in the first
internal NO
2
to NO converter.
Conv. Efficiency NH3
The overall efficiency of the conversion of NH
3
to NO, in the External
converter, combined with the efficiency of the second internal NO
2
to
NO converter.
Noise
The standard deviation of the concentration. The calculation is as
follows:
Take a concentration value once every two minutes.
Store 25 of these samples in a first in, last out buffer.
Every two minutes, calculate the standard deviation of the current

Table of Contents

Related product manuals