EasyManua.ls Logo

Ecotech Serinus 44 - 4.7.3 Instrument Control

Ecotech Serinus 44
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Communications
Page 85
5. A screen shot of the instrument’s current screen should appear on your Smartphone or
tablet. To disconnect press the back key/button on the device.
Note: Once the instrument has been paired with the device it will appear under “Paired Devices” and
the PIN will not need to be entered again. Only one Bluetooth connection can be made to an
instrument at any one time.
4.7.3 Instrument Control
Once connected the user has full control of the instrument. The range for remote control depends on
the device’s Bluetooth capabilities and any intervening obstructions, but is usually up to 30 meters.
Remote Screen Operation
With the exception of the number pad, all button functions/actions can be performed by touching
the screen. This includes the selection buttons and the scroll buttons. Touching any part of the
screen where there is not already a button also enacts the functions of the scroll buttons.
Home Screen
Touching the upper half of the screen increases the contrast and touching the lower half of the
screen decreases contrast on the real instrument.
Menus
Touching the upper or lower half of the screen allows the user to scroll up and down respectively.
Right-hand Section of the Screen
Swiping from right to left brings up the number pad for entering numbers (swipe from left to right to
hide the number pad).
Figure 38 Showing or Hiding the NumPad
Left-hand Section of the Screen
Swiping from left to right brings up a list of available analysers (swipe from right to left to hide the
instrument list).

Table of Contents