Communications
Page 73
Figure 20 External 25 Pin I/O Individual Pin Descriptions
CAUTION
The analog and digital inputs and outputs are rated to CAT I.
Exceeding 12VDC or drawing greater than 400mA on a single output or a total
greater than 2A across the eight outputs can permanently damage the instrument
and void the warranty.
4.5 Logging Data
When the user receives the instrument from the factory it will have a default set of parameters
already setup in the internal data logger. These select few parameters have been chosen for their
relevance in assisting in troubleshooting the instrument.
4.5.1 Configure Instrument Internal Logging
In order to log data you must first specify a data logging interval. This is how often data will be logged
to the USB memory stick. You can log a maximum of 12 parameters. These parameters can be
selected by name or by parameter number using Appendix A (refer to Table 12).
Procedure
1. Open - Main Menu Communications Menu Data Logging Menu (refer to Section 3.4.21).
2. Select - Data Log Interval (adjust to the desired value).