Regulations generally require that the instrument be span calibrated against a certified calibration 
source any time: 
  The instrument is moved. 
  The instrument is serviced. 
  The instrument’s units are changed by the user between volumetric and gravimetric. 
  Whenever the instrument characteristics may have changed. 
Regulatory agencies establish the time intervals at which the instrument must be calibrated to 
ensure satisfactory data for their purposes. 
5.2  Pressure Calibration  
The pressure sensors are a vital component of the instrument operation and the pressure calibration 
should be checked on installation or whenever maintenance is performed. 
A thorough leak check must be performed prior to performing a pressure calibration (refer to Section 
6.3.5). 
The pressure calibration can either be a two point calibration (one point under vacuum and another 
point at ambient pressure) or a single ambient point calibration when very minor adjustments are 
required. 
Note: Ensure that the instrument has been running for at least one hour before any calibration is 
performed to ensure the instrument’s stability. When performing a two point pressure calibration, it 
is advisable to perform the vacuum pressure calibration first. 
 
5.2.1  Full Pressure Calibration  
This section outlines how to perform a full pressure calibration. Using the required equipment follow 
the steps below to complete a full pressure calibration.  
Note: This procedure is for firmware version 3.17.0007 and above. Ensure that the instrument has 
been running for at least one hour before the calibration is performed.  
 
Note: Ensure units of measure are the same on both the barometer and instrument. 
Equipment Required 
  Barometer 
  Vacuum source 
Procedure 
1.  Disconnect all external tubing connected to the rear ports of the instrument except for the AUX 
in Port and AUX Out Port (for the external converter). 
2.  Open - Main Menu  Calibration Menu  Pressure Calibration Menu - (read note) - OK.