Serinus 44 User Manual 3.0
Page 18
7 days: < 1.0% of reading
1.2.3 Power
Operating voltage
100-240VAC, 50 to 60 Hz (Autoranging)
Power Consumption
265 VA max (typical at start-up)
190 VA after warm-up
1.2.4 Operating Conditions
Ambient Temperature Range
0 °C to 40 °C (32 °F to 104 °F)
Sample Pressure Dependence
5% change in pressure produces less than a 1% change in reading
Maximum altitude: 3000m above sea level
1.2.5 Communications
Analog Output
Menu selectable current output of 0-20 mA, 2-20mA or 4-20mA.
Voltage output of 0 to 5 V, with menu selectable zero offset of 0%, 5%, or 10%.
Voltage output of 0 to 10 V (configured using jumpers (JP3) on rear panel PCA).
Range: 0 to full scale from 0-0.05 ppm to 0-20 ppm with 0%, 5%, and 10% offset.
Analog Input
Three analog voltage inputs (0-5VDC) CAT I rated.
Digital Output
RS232 port #1: Normal digital communication or termination panel connections.
RS232 port #2: Multidrop port used for multiple instrument connections on a single RS232.
USB port connection on rear panel.
25 pin connector with discrete status and user control.
USB memory stick (front panel) for data logging, event logging, parameter and configuration
storage.
Eight Digital Outputs, open collector max 400mA each @ 12VDC (max total output 2A).
Eight Digital Inputs, 0-5VDC, CAT I rated.