Serinus 44  User Manual  3.0 
 
 
7.7  Ecotech Service Support Files 
Regular backup of the settings, parameters and data on the instruments USB memory stick is 
recommended. 
In the event of a fault that requires Ecotech technical support, please make copies of the following 
files and email to: support@ecotech.com 
Equipment Required 
  PC/Laptop 
Procedure 
State the ID number, variant, board revision and firmware version of the instrument with a brief 
description of the problem. Take a copy of the current configuration if possible and a save of the 
parameters. 
1.  Open - Main Menu  Analyser State Menu. 
2.  Variant - (take note). 
3.  Ecotech ID - (take note). 
4.  Board Revision - (take note). 
5.  Firmware Ver. - (take note). 
6.  Open - Main Menu  Service Menu. 
7.  Save - Save Configuration - (CONFIG**.CFG) - Accept. 
 
 
8.  Save - Save Parameter List - (PARAM**.TXT) - Accept. 
9.  Eject - Safely Remove USB Stick - (Follow instructions). 
 
Note: PARAM99.TXT is the “Factory Backup” file, This is a snap shot of the parameters while it was 
under test in the factory just prior to release.  It is recommended that this file is kept unchanged but 
can be viewed for reference.  
** Can be any number from 0 – 98. 
Note: CONFIG99.CFG is the “Factory Backup” file, this is the configuration of the instrument as it left 
the factory. It is recommended that this file is kept unchanged but can be used as a reference backup 
point. 
 ** Can be any number from 0 – 98.