Figure 65 – Cleaning of Glass Tube
6. With the PCA now free locate the white Teflon block fastening screws and disassemble glass tube
(refer to Section 9.7).
7. Clean or replace the glass tube and replace the Teflon gaskets.
8. Reassemble in the reverse order.
9. Before installing into the instrument perform a leak check to ensure your ozone generator
assembly is not leaking.
10. Install the ozone generator assembly into the instrument. Power up the instrument and vacuum
pump and wait for the warm-up sequence to complete.
11. Perform a full system leak check (refer to Section 6.3.5).
12. Perform a zero and span calibration.