Serinus 44  User Manual  3.0 
 
 
3.4.14  Digital Pots Menu 
Main Menu  Service Menu  Diagnostics Menu  Digital Pots Menu 
Digital pots are electronically controlled digital potentiometers used for adjustments to operations of 
the instrument. Each of the digital pots can go from 0-255. This menu should be accessed only during 
diagnostics. 
Unless the Control Loop is Disabled (refer to Section 3.4.13), changes to the pots may be modified by 
the instrument. This is intentional; some diagnostics are best done with instrument feedback and 
some are best done without. 
When the instrument first starts it will tune the high voltage 
supply by automatically setting the High Volt Adj Pot. After a 
stable value is reached the instrument will Disable the High Volt 
Tuning. You can force the instrument to re-tune the high voltage 
supply by setting this field to Enabled and rebooting the 
instrument. 
Allows manual adjustment of the PMT high voltage supply. 
The voltage applied to the PMT. 
Displays the gain of the PGA. 
Reduces the raw signal to measurable level. 
The concentration voltage measured by the analog to digital 
converter. 
The concentration voltage after adjustment for the PGA gain 
factor. 
  Operate (default): Puts the instrument in normal operation 
mode. 
  Electrical: Injects an artificial test signal into the electronic 
processing circuitry on the main controller PCA to verify that 
the circuitry is operating correctly. When in this Diagnostic 
Mode, adjust the Diagnostic Test Pot from 0 to 255. This will 
produce a change in the concentration voltage as well as the 
indicated gas concentration.  
  Preamp: Injects an artificial test signal into the Preamplifier 
module mounted in the optical bench to verify that the 
Preamplifier, cabling and electronic circuitry on the main 
controller PCA is operating correctly. When in this Diagnostic 
Mode, adjust the Diagnostic Test Pot from 0 to 255. This will 
produce a change in the concentration voltage as well as the 
indicated gas concentration.  
  Optic [Optional]: Emits artificial light into the Reaction Cell to 
simulate a real chemiluminescence reaction. This will verify that 
the PMT, Preamp and electronic circuitry on the main controller 
PCA is operating correctly. When in this Diagnostic Mode, 
adjust the Diagnostic Test Pot from 0 to 255. This will produce a 
change in the concentration voltage as well as the indicated gas 
concentration. This menu item is only available if you have 
installed the optical test lamp (refer to Section 3.4.30). 
This Digital Pot is used for diagnostics only. When in the Electrical,