EasyManua.ls Logo

Ecotech Serinus 44 - Figure 28 - New Station Connection; Table 5 - New Station Setup

Ecotech Serinus 44
174 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Communications
Page 79
Figure 28 New Station Connection
Table 5 New Station Setup
Property
Description
Station Name
The name of the station. If you have other loggers, the name is used to distinguish them.
Logger
Set this to “Ecotech Serinus” when downloading from any Serinus series instrument. This
will communicate with the instrument via the Advanced protocol. If using a network or
serial connection, ensure that the Advanced protocol has also been selected on the
instrument itself.
Time Zone
Set this to the time zone that the instrument is located in.
DST
Enable this option if you plan on changing the clock on the instrument with daylight
savings. Leave this disabled if the clock does not shift during DST. The instrument will need
to be adjusted manually for DST it will not happen automatically.
Database Name
This is the name to be used for the table in the SQL database containing this station’s data.
It must be unique for each station.
Device ID
Enter the Serial ID of the instrument. If you are not using multidrop; this can be set to “0”
or left blank.
Link Type
Select the type of connection used to connect to the instrument. Different properties will
appear depending on the link type selected. Align these settings with those of the
instrument.
Log Interval
This needs to be the same as the Data Log Interval setting on the instrument.
Note: The available fields for communication parameters will change when you change the link type.
You will need to set the communication parameters that have been defined on the instrument.
5. Once the station has been created, save the station by clicking the Save shortcut icon or
File Save.

Table of Contents

Related product manuals