Serinus 44 User Manual 3.0
Page 170
<STX>{DA}<ETX>< bcc1><bcc2>
Where:
kkk = Device’s Serial ID
bcc1 = First byte of the block check calculation
bcc2 = Second byte of the block check calculation
Device response (S10, S30 and S50 family)
<STX>{MD}{01}<SP><kkk><SP><+nnnn+ee><SP><ss><SP><ff><{000}><SP>{00000000}
<SP><ETC>< bcc1><bcc2>
Device response (S40 family)
<STX>{MD}{02}<SP><kkk><SP><+nnnn+ee><SP><ss><SP><ff><SP>{00000000}
<SP><mmm><SP><+pppp+ee><SP><ss><SP><ff><SP>{00000000}
<SP><ETC><bcc1><bcc2>
Where:
<SP> = Space (0x20 hex)
kkk = Device’s Serial ID. If the DA command is issued without an ID, the response omits this field.
Exception: the S40 family always includes both ID fields, even when a DA command without an ID is
issued.
+nnnn+ee = Main instantaneous gas concentration (for S40 family, this is NO)
ss = Status byte with the following bit map:
Table 16 Bit Map
Status Bit
Meaning if set to 1
0
Instrument off (this value is always set to 0).
1
Out of service.
2
Zero mode.
3
Span mode.
4
-
5
-
6
Units: 1 = Volumetric, 0 = Gravimetric.
7
Background mode (S30 and S50 family only).
ff = Failure byte for both channels with the following bit map (positive logic):