Operation
Page 63
CAUTION
No items in this menu should be accessed without authorization and supervision
of qualified service personnel.
3.4.30 Hardware Menu
Advanced Menu Hardware Menu
This menu contains factory hardware installations. If you reset to factory defaults you may need to
revisit this menu to re-install optional features.
Variant
Select the instrument model. Normally this only needs to be reset
when the configuration is corrupted. The selections available will
depend on licensing. It is not recommended to run an instrument
with firmware set to an incorrect model.
Range
Select the range of the instrument. Not all instruments support all
ranges. Default is Standard.
Front Panel Style
Choosing the incorrect front panel will result in the traffic lights
behaving inconsistently. Default is Aluminium.
Network Port
When Enabled indicates the instrument has a network port
installed. Default is Disabled.
Orifice Size
Specify the input orifice. Default is 0.96.
Optical Test Lamp
When Enabled gives access to the optical diagnostic tests. Default
is Disabled.
CAUTION
No items in this menu should be accessed without authorization and supervision
of qualified service personnel.
3.4.31 Parameter Display Menu
Main Menu Advanced Menu Parameter Display Menu
Used to display a parameter in real-time on the screen (refer to Table 12 for a full list of parameters).
Data Parameter
This is an editable field. Enter the parameter number you wish to
view (refer to Table 12)
Name
Displays the name of the selected parameter.
Value
Displays the current value of the selected parameter.