Operation
Page 51
3.4.12 Service Menu
Main Menu Service Menu
Diagnostics Menu
Refer to Section 3.4.13.
Calculation Factors Menu
Refer to Section 3.4.19.
Ozone Gen. Control
Allows the user to switch off the ozone generator or let the instrument
automatically control the status of the ozone generator. For safety
reasons, the ozone generator will not be active if there is a flow fault.
Ozone Gen. Status
This field indicates to the user if the Ozone generator is On or Off.
Load Auto-Backup Config.
Loads the auto-backup configuration file. The configuration is
automatically backed up every night at midnight.
Load Configuration
Loads a user selectable configuration file from the USB memory stick.
Save Configuration
Saves all of the EEPROM-stored user-selectable instrument
configurations to the USB memory stick (calibration and communication
settings, units, instrument gain, etc). If there are problems with the
instrument use this function to save settings to the USB memory stick
and send this file (together with the parameter list save) to your
supplier with your service enquiry.
Save Parameter List
Saves a text file of various parameters and calculation factors. If you
have problems with the instrument use this function to save settings to
the USB memory stick and send this file (together with the
configuration save) to your supplier with your service enquiry.
Instrument
This field allows the instrument to be set to either Online (normal
instrument operation) or In Maintenance (data is flagged as invalid).
Next Service Due
Displays when the next scheduled service is due.
Safely Remove USB Stick
This command must be activated to safely remove the USB memory
stick.
System Restart
Activating this will restart the instrument.
3.4.13 Diagnostics Menu
Main Menu Service Menu Diagnostics Menu
Digital Pots Menu
Refer to Section 3.4.14.
Valve Menu
Refer to Section 3.4.15.
Tests Menu
Refer to Section 3.4.16.
Pres/Temp/Flow Comp.
On (default): Is used to compensate instrument measurements for
environmental fluctuations that might affect readings (pressure,
temperature and flow).
Off: Is used only when running diagnostics.
Control Loop
Enabled (default): Allows the instrument to automatically adjust digital
pots and other outputs.
Disabled: Prevents the instrument from changing most outputs so the
service technician can manually control them.