Operation
Page 45
power was disconnected from the instrument.
3.4.4 Status Menu
Main Menu Analyser State Menu Status Menu
The Status Menu presents a list of the current Pass/Fail statuses of the main components.
During warm-up, the status of some parameters will be a dashed line.
Event Log
This field enters a screen with a log of all the events that the instrument
has performed. These events include errors and warnings. This log is
stored on the USB memory stick.
The log is organised by month. When you enter this screen you will be
prompted to enter the month for which you wish to view events.
Show Error List
This field allows the user to display the list of current errors and
warnings on the screen.
Next Service Due
This field is visible with the next service due date if the service is due
within the next two weeks.
+5V Supply
Pass if the +5V power supply is within the acceptable range.
+12V Supply
Pass if the +12V power supply is within the acceptable range.
+Analog Supply
Pass if the analog power supply is within the acceptable range (+12V).
-Analog Supply
Pass if the analog power supply is within the acceptable range (-12V).
A2D
Fail only if a problem is detected with the analog to digital conversion.
Cell Temp.
Pass if the cell heater temperature is within ±10% of the heater set
point (refer to Section 3.4.5).
Converter Temp.
The converter temperature should be 325°C. It must be between 250°C
and 350°C to pass.
Cooler
Status of the PMT cooler. It must be 13°C ±10% to pass.
High Voltage
Fail if the high voltage value is <20 or >30 from the target.
Target is 650V (Standard), 700 (Trace) and 550(High).
System Power
Pass if the system has an adequate electrical supply.
Maintenance Mode
Error if the system is “In Maintenance (refer to Section 3.4.12).
Diagnostic Mode
Error if the electronics are in Diagnostic Mode (refer to Section 3.4.14).
Diagnostic PTF Comp
Error if the Pres/Temp/Flow Comp. is disabled (refer to Section 3.4.13).
Diagnostic Control
Error if the control loop is disabled (refer to Section 3.4.13).
Valve Manual Control
Error if the valve sequencing is disabled (refer to Section 3.4.15).
Ozonator
Indicates if the Ozonator is on or off.
NO Conc V Saturated
Indicates if the voltage of the concentration during measurement is
within the limits of the analog to digital converter (-0.26 to 3.29V).
NOx Conc V Saturated
Indicates if the voltage of the concentration during measurement is
within the limits of the analog to digital converter (-0.26 to 3.29V).
Nx Conc V Saturated
Indicates if the voltage of the concentration during measurement is
within the limits of the analog to digital converter (-0.26 to 3.29V).