Serinus 44 User Manual 3.0
5. The same procedure can be followed to also activate zero calibrations. Pin 6 of the instrument’s
analog/digital 25 pin female connector can be connected to one of the other master devices
digital outputs and the instrument can be set so DI 1 is assigned to Do Zero.
3.4.26 Digital Output Menu
Main Menu Communications Menu Digital Output Menu
This allows the instrument to trigger external alarms in response to certain events.
There are eight different pins available, which will be set high during an associated event:
Associates a state with a digital output. There are eight digital
outputs (the pin numbers are for the 25 pin connector). Each one
can have one of the associated states listed in Table 3 – Digital
Output States. The pin will be driven to the active state while the
instrument state is true.
Each pin can be set to be active High or Low. Active High means
that the pin will be pulled to 5V when the associated event occurs.
Active Low means the pin will be pulled to 0V when the associated
event occurs.
Table 3 – Digital Output States
No state (this state is never Active).
Chassis temperature fault.
Cooler temperature fault.
The USB memory stick is disconnected.
Any system fault (the red light is on).