mode is set in Cycle Time (refer to Section 3.4.10).
While the instrument is still in the warm-up period (refer to Section 3.1)
the Cal. Mode cannot be changed from Measure mode.
3.4.10.2 Timed Mode
These items appear in the Calibration Menu when Calibration Type is set to Timed.
Date
Enter the date for the next calibration to start.
Time
Enter the time that calibration will be performed. The time is set using a
24 hour clock.
Repeat
This field indicates the delay period; once the specified amount of time
has lapsed the calibration will automatically run again. The user can edit
this field (from 1 to 20,000 units).
Units
This is where the user can define the type of units for the Repeat delay
period. For example: A repeat of 3” and units of “Days means that a
calibration will automatically be performed every three days.
Span Compensation
When Enabled the instrument will automatically perform a Span
Calibrate NOx at the end of the Cycle Time and adjust the gain based
on the Span Level.
When Disabled it will do a precision check only, no adjustment is made.
Span Level
Enter the concentration of span gas expected. Used when the Span
Compensation is Enabled.
3.4.11 Pressure Calibration Menu
Main Menu Calibration Menu Pressure Calibration Menu
Entering this menu will set the valves to the pressure calibration configuration and will disable the
ozone generator; leaving the menu will restore the valves to normal operation and enable the ozone
generator (refer to Section 5.2).
Vacuum Set Pt.
The zero point for the calibration. Activating this item will open a dialog
box of instructions.
Ambient Set Pt.
The high point for the calibration. Activating this item will open a dialog
box of instructions.
Pressure Units
Select the units that the pressure will be displayed in (torr, PSI, mBar,
ATM or kPa).
Ambient
The current ambient pressure.
The current ambient pressure displayed as a raw voltage.
Cell
The current pressure in the reaction cell.
The current reaction cell pressure displayed as a raw voltage.
Manifold
The current pressure in the auxiliary valve manifold.
The current auxiliary valve manifold pressure displayed as a raw
voltage.