port or from the internal zero source when zero gas is requested.
The duration of each Cal. Mode (span and zero) when performing Cycle
Mode (refer to Section 3.4.10.1) or Cal. Type is set to Timed (refer to
Section 3.4.10.2).
This field is used to perform a span calibration and should be only used
when a known concentration of span gas is running through the
reaction cell and the reading is stable.
Activating the span calibrate field for a named gas will open a dialog
box. Enter the concentration of the span gas that the instrument is
sampling and press Accept.
This command is used to correct the zero calibration setting. This option
should be used only when zero gas is running through the reaction cell
(refer to Section 5.3 before using this command).
This command is used to correct the zero calibration setting. This option
should be used only when zero gas is running through the reaction cell
(refer to Section 5.3 before using this command).
This command is used to correct the zero calibration setting. This option
should be used only when zero gas is running through the reaction cell
(refer to Section 5.3 before using this command).
Pressure Calibration Menu
This menu allows the user to calibrate the pressure sensors (refer to
Section 3.4.11).
This field displays the auxiliary valve manifold pressure measured
during the last calibration for the NO path.
This field displays the auxiliary valve manifold pressure measured
during the last calibration for the NO
X
path.
This field displays the auxiliary valve manifold pressure measured
during the last calibration for the N
X
path.
Cell temperature when the last span calibration was performed.
The overall efficiency of the conversion of NO
2
to NO, in the first
internal NO
2
to NO converter.
The overall efficiency of the conversion of NH
3
to NO. Taking into
account the combined efficiency of both the External NH
3
converter and
the second internal NO
2
to NO converter.
3.4.10.1 Manual Mode
These items appear in the Calibration Menu when Calibration Type is set to Manual.
When calibration type is set to Manual the instruments operational
mode can be chosen from the following:
Measure: Is the normal measurement through the sample port.
Zero: This mode will take air through the calibration port so that a
zero calibration can be performed. Data is flagged as zero data.
Span: This mode will take air through the calibration port so that a
span calibration can be performed. Data is flagged as span data.
Cycle: Performs a zero and a span Cal. Mode and then returns to
measure mode. The length of time spent measuring each calibration