Serinus 44 User Manual 3.0
Page 62
3.4.28 Bluetooth Menu
Main Menu Communications Menu Bluetooth Menu
This instrument supports Bluetooth communication through the Serinus Remote Android Application
(refer to Section 4.7).
Bluetooth
This field indicates whether the instrument is remotely connected
to an Android device.
Reset Bluetooth
After changing the ID or PIN, it is necessary to reboot the
Bluetooth module. This is done by resetting the instrument or by
using this menu item to reboot only the Bluetooth.
ID
This is the Bluetooth ID of the instrument. Use the keypad to edit
this field (refer to Section 3.3.1 for instructions on entering text
with the numeric keypad).
The default ID setting is Serinus(Ecotech ID).
The word Serinus is always the first part of the name and cannot
be edited. The second part is the Ecotech ID.
PIN
This is a passcode/pin required for the Serinus Remote Application
to connect to the instrument. The default pin is 1234.
3.4.29 Advanced Menu
This menu is accessed via a different method than the other menus. From the Home Screen press
the following keys: ()99(

)
This menu contains technical settings, diagnostics and factory hardware installations. No items in this
menu should be accessed without authorisation and supervision of qualified service personnel.
Language
Select a language.
Hardware Menu
Refer to Section 3.4.30.
Service Displays
When set to On, new items appear on many different menus.
These fields are for diagnostic and service personnel only. Default
is Off.
Next Service Due
Enables the user to edit the next service due date.
Jump to Next State
Moves the sequence to the next state (e.g. from Fill to Measure).
This command is most commonly used to force an instrument out
of the warm-up sequence early.
Parameter Display Menu
Refer to Section 3.4.31.
Reset to Factory Defaults
Reset the configuration to factory defaults. This will erase all
calibrations and user configuration information.
Rebuild Index
If a data log becomes corrupted it may be possible to restore It by
rebuilding its index file. This command will ask you to specify a
month and will rebuild the index for that month. This operation
can take many minutes and it should not be interrupted. While the
file is rebuilding any data logging will be suspended.