Digital Status Outputs
The instrument is equipped with eight open collector outputs, which will convey instrument status
conditions and warning alarms such as no flow, sample mode, etc.
CAUTION
Exceeding 12VDC or drawing greater than 400mA on a single output or a total
greater than 2A across the eight outputs can permanently damage the instrument
and void the warranty.
Bluetooth
This allows for remote access of the instrument to any Android device with the Serinus remote
application installed. It uses Bluetooth to control the instrument, view parameters, download data
and construct real-time graphs.