SCU-800 Control Unit for Turbomolecular Pump
*1
The driver output stops and the rotor continues rotating by
inertia.
It may take several hours to stop. To stop the STP pump quickly, close the vacuum valve at the outlet port
flange and introduce gas from the purge port into the STP pump.
Table
39 - Error
messages after self test (continued
)
1. Contact
Service office
.
2. Contact
Service office
.
Overhaul is needed.
Contact
Service office
.
Set "Warning Damage Point"
to "DISABLE" to release the
error and continue operation
temporary.
1. Connect the cable
correctly. If the cable
connecto
r is not locked,
lock it securely.
2. Contact
Service office
.
3. When the TMS function is
not in use, set the TMS
function to "DISABLE"
.
1. Connect the
electromagnetic cooling
water valve correctly.
Check the volume of water
and water temperature.
2. Contact
Service
office
.
2. Failure of the SCU
-800.
Frequency of "Disturbance"
or
"Mains
Failure
" error.
1. Disconnection of the TMS
connection cable.
2. Failure of the TMS
connection cable.
3. Incorrect setting of the
TMS function.
1. Cooling fail
ure of
TMS unit
.
Abnormal rotational speed
detection.
"Second Damage Limit"
occurs.
TMS connection cable is
not connected when the
state "TMS Function" is
set to "ENABLE"
.
TMS temperature control
error (when the TMS unit
detection temperature
becomes higher than the
setting value by 10°C (+18
ºF) or more.