EasyManua.ls Logo

Edwards SCU-800

Edwards SCU-800
141 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
SCU-800 Control Unit for Turbomolecular Pump
MT-49E-101-G
Page 120
7
MAINTENANCE
Referred
section
Instruction
manual
(B)
3.11
.3
3.6.2
3.9
.1
Table
40 - Error
messages after self test (continued)
Countermeasures
1. Contact Service office.
2. Contact Service office.
1. Contact Service of
fice.
2. When the TMS function is
not in use, set the TMS
function to "DISABLE"
.
3. Connect the cable
correctly. If the cable
connector is not locked,
lock it securely.
4. Contact Service office.
1. Perform tuning after
power
ON/OFF
operation.
2. When the error is displ
ayed
after tuning, contact
Service office.
Probable causes
1. Failure of the TMS heater.
2. Failure of the TMS unit.
1. Disconnection of the TMS
sensor.
2. Incorrect setting of the
TMS function.
3. Disconnection of the TMS
connection cable.
4. Failure of the TMS
connec
tion cable.
1. Failure of the STP
pump.
2. Failure of the SCU
-800.
Pump operation
Decelerate and stop.
Decelerate and stop.
Touch down. The
magnetic bearing
output stops and the
rotor is descended
on the
touch down
bearing.
Occurrence condition
TMS temperat
ure control
error (when the TMS unit
detection temperature
becomes lower than the
setting value by 10°C
(-18
ºF) or more.
Error is not
detected for a certain time
after the STP pump starts.
TMS sensor is not
connected when the state
"TMS Function" is
"ENAB
LE"
.
Tuning failure.
LCD Messages
TMS Lower Temp
TMS Sensor Lost
Tuning Error 1
:
:
Tuning Error
19

Table of Contents

Related product manuals