STP-iXR1606 Turbomolecular Pump
MT-79E-001-D
Page 101
5
SERIAL COMMUNICATION PROTOCOL
5.4.6 ReadVersion
Function: Read the software version.
Transmission frame:
PC->SIM Stx 0 0 1 ? V Etx LRC
SIM->PC Ack
PC->SIM Ack
SIM->PC Stx 0 0 1 Sp V Parameter 1 to 24 Etx LRC
Parameter Item Data format Remarks
1 to 16 Control unit software version 8-bits hexadecimal
coded ASCII
17 to 20 Motor driver software version 4-bits ASCII character Ver.1.2 = 0120
21 to 24 AMB software version 4-bits ASCII character Ver.3.4 = 0340
Example:
Control unit software version : 37395F4120312E302020202020202020hex = 79_A 1.0
Motor driver software version : 0120
= 1.2
DSP software version : 0340
= 3.4
Parameter 1 2 3 4 5 6 7 8 9 10
"7 "9” "_" "A" " " "1" "." "0" " " " "
ASCII 3 7 3 9 5 F 4 1 2 0 3 1 2 E 3 0 2 0 2 0
HEX 33 37 33 39 35 46 34 31 32 30 33 31 32 45 33 30 32 30 32 30
Parameter 11 12 13 14 15 16 17 18 19 20 21 22 23 24
" " " " " " " " " " " "
ASCII 2 0 2 0 2 0 2 0 2 0 2 0 0 1 2 0 0 3 4 0
HEX 32 30 32 30 32 30 32 30 32 30 32 30 30 31 32 30 30 33 34 30