STP-iXR1606 Turbomolecular Pump
MT-79E-001-D
Page 106
5
SERIAL COMMUNICATION PROTOCOL
Parameter Item Data format Remarks
1 The number of
"Error Record"
8-bits hexadecimal
coded ASCII
Up to 10 errors
2 to 11 Error Record 1 to
Error Record 10
8-bits hexadecimal
coded ASCII
*1
*1 Value corresponding to the error message is transmitted (refer to Table 28 and Table 29). The most recent error has the
smallest parameter number. When the number of errors being detected is under the maximum number, the value of
parameter that is larger than the number of errors has been detected is set to 0.
Example:
When 3 errors have been detected in the past;
The number of "Error Record" : 03
hex
= 3 errors
Error Record 1 : 0F
hex
= 15 = Disturbance Xb
Error Record 2 : 0D
hex
= 13 = Disturbance Xh
Error Record 3 : 12
hex
= 18 = T.Cable Disconnected
Error Record 4 to 10 : 00
hex
= No error recorded
Parameter 1 2 3 4 5 6 7 8 9 10 11
ASCII 0 3 0 F 0 D 1 2 0 0 0 0 0 0 0 0 0 0 0 0 0 0
HEX 30 33 30 46 30 44 31 32 30 30 30 30 30 30 30 30 30 30 30 30 30 30
5.4.12 SetSpeedSetPoint
Function: Changes the "Speed Set Point" value. This value can be changed the range from 18,500 to
36,500 rpm. The threshold value of the illumination pattern of the "ROTATION" LED is fixed. It is not
changed even if the setting value of the rotational speed is changed.
Transmission frame:
PC->SIM Stx 0 0 1 Sp h Parameter 1 Etx LRC
SIM->PC Ack
PC->SIM Ack
SIM->PC Stx 0 0 1 # Etx LRC