EasyManua.ls Logo

Edwards STP Series

Edwards STP Series
169 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
STP-iXR1606 Turbomolecular Pump
MT-79E-001-D
Page 116
5
SERIAL COMMUNICATION PROTOCOL
5.4.17 SetOptionFunc
Function: Changes the setting value of the input operation port, warning function, and serial
communication time out.
Transmission frame:
*1
PC->SIM Stx 0 0 1 Sp = Parameter 1 to 13 Etx LRC
SIM->PC Ack
PC->SIM Ack
SIM->PC Stx 0 0 1 # Etx LRC
*1 The HEX code of ASCII character ' = ' is "3D".
Parameter Item Data format Remarks
1 Input operation port setting 8-bits hexadecimal coded
ASCII
Default 01
hex
(I/O REMOTE)
2
*
1
TMS Option
Enable/Disable setting
8-bits hexadecimal coded
ASCII
Default FF
hex
(DISABLE)
3 [System reservation] 48-bits hexadecimal
coded ASCII
*
2
4 Second Damage Limit Option
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default 00
hex
(ENABLE)
5 First Damage Limit Warning
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default 00
hex
(ENABLE)
6
Pump Runtime Over Warning
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default FF
hex
(DISABLE)
7 Pump Runtime Over Warning
Hours setting (×100 hours)
32-bits hexadecimal
coded ASCII
Default
00000000
hex
8
Imbalance Warning
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default 00
hex
(ENABLE)
9 Pump Overload Warning
Enable/Disable setting
8-bits hexadecimal
coded ASCII
Default FF
hex
(DISABLE)
10
Pump Overload Warning
Motor current setting
16-bits hexadecimal
coded ASCII
Default 03E8
hex
(100.0%)
11
Pump Overload Warning
Rotational speed setting
16-bits hexadecimal
coded ASCII
Default 0000
hex
(0.0%)
12 Serial communication
time out setting (Unit: sec.)
16-bits hexadecimal
coded ASCII
Default 003C
hex
(60 sec.)

Table of Contents