Page #128 background imageLoading...
Page #128 background image
STP-iXR1606 Turbomolecular Pump
MT-79E-001-D
Page 118
5
SERIAL COMMUNICATION PROTOCOL
5.4.18 ReadCondition
Function: Reads the pump model and damage point.
Transmission frame:
*1
PCSIM
Stx 0 0 1 ? { Etx LRC
SIMPC
Ack
PCSIM
*1
Ack
SIMPC
Stx 0 0 1 Sp { Parameter 1 to 4 Etx LRC
*1 The HEX code of ASCII character ' { ' is "7B".
Parameter Item Data format Remarks
1 Pump model 160-bits hexadecimal
coded ASCII
2 [System reservation] 32-bits hexadecimal
coded ASCII
3 Damage point 16-bits hexadecimal
coded ASCII
4 [System reservation] 64-bits hexadecimal
coded ASCII

Table of Contents