Installation Cubemass DCI MODBUS RS485
12 Endress+Hauser
3.2.2 Mounting location
The accumulation of air and the formation of gas bubbles in the measuring tube could result in an
increase in measuring errors.
For this reason, avoid the following mounting locations in the pipe:
The highest point of a pipeline. Risk of air accumulating!
Directly upstream of a free pipe outlet in a down pipe.
A0003605
Fig. 5: Mounting location
The proposed configuration in the following diagram, however, permits installation in an open
down pipe. Pipe restrictors or the use of an orifice plate with a smaller cross-section than the
nominal diameter prevent the sensor from running empty during measurement.
A0003597
Fig. 6: Installation in a down pipe (e.g. for batching applications)
1 Supply tank
2Sensor
3 Orifice plate, pipe restriction (see Table)
4Valve
5 Batching tank
1
2
3
4
5
DN Orifice plate, pipe restriction
mm inch mm inch
11/24"0.80.03
21/12"1.50.06
4 1/8" 3.0 0.12
6 ¼" 5.0 0.20