Installation Cubemass DCI MODBUS RS485
16 Endress+Hauser
3.3.2 Installing the wall-mount housing
There are various ways of installing the wall-mount housing:
• Mounted directly on the wall
• Installation in control panel (separate mounting set, accessories) → ä 17
• Pipe mounting (separate mounting set, accessories) → ä 17
"
Caution!
• At the mounting location, ensure that the permitted ambient temperature range
(–20 to +60 °C (–4 to +140 °F), optional –40 to +60 °C (–40 to +140 °F)) is not exceeded.
Install the device in a shady location. Avoid direct sunlight.
• Always install the wall-mount housing in such a way that the cable entries are pointing down.
Mounted directly on the wall
1. Drill the holes → å 9.
2. Remove the cover of the connection compartment (a).
3. Push the two securing screws (b) through the appropriate bores (c) in the housing.
– Securing screws (M6): max. Ø 6.5 mm (0.26")
– Screw head: max. Ø 10.5 mm (0.41")
4. Secure the transmitter housing to the wall as indicated.
5. Screw the cover of the connection compartment (a) firmly onto the housing.
A0001130
Fig. 9: Mounted directly on the wall
a
b
c
c
90 (3.54)
35 (1.38)
192 (7.56)
81.5 (3.2)
mm (inch)