Commissioning Cubemass DCI MODBUS RS485
58 Endress+Hauser
6.4.4 Pulse/frequency output
The configuration of the pulse/frequency output with line monitoring "On" or "Off" takes place by
means of various jumpers on the pulse/frequency output submodule.
#
Warning!
Risk of electric shock. Exposed components carry dangerous voltages.
Make sure that the power supply is switched off before you remove the cover of the electronics
compartment.
1. Switch off power supply.
2. Remove the I/O board → ä 77.
3. Set the jumpers (→ å 35).
"
Caution!
Risk of destroying the measuring device. Set the jumpers exactly as shown in the diagram.
Incorrectly set jumpers can cause overcurrents that would destroy either the measuring device
or external devices connected to it.
4. Installation of the I/O board is the reverse of the removal procedure.
A0011939
Fig. 35: Configuring pulse/frequency outputs with the aid of jumpers (I/O board)
1 Pulse/frequency output 1
1.1 Line monitoring ON (factory setting)
1.2 Line monitoring OFF
1.1
1.2
+
+