EasyManua.ls Logo

Endress+Hauser Cubemass DCI - 6.5 Adjustment

Endress+Hauser Cubemass DCI
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Commissioning Cubemass DCI MODBUS RS485
60 Endress+Hauser
6.5 Adjustment
6.5.1 Zero point adjustment
All measuring devices are calibrated with state-of-the-art technology.
The zero point obtained in this way is printed on the nameplate.
Calibration takes place under reference operating conditions ä 86.
Consequently zero point adjustment is generally not necessary for Cubemass DCI!
Experience shows that the zero point adjustment is advisable only in special cases:
To achieve highest measuring accuracy also with very small flow rates.
Under extreme process or operating conditions (e.g. very high process temperatures).
Preconditions for a zero point adjustment
Note the following before you perform a zero point adjustment:
Adjustment can only be performed on homogeneous fluids.
Zero point adjustment is performed at zero flow (v = 0 m/s). This can be achieved, for example,
with shutoff valves upstream and/or downstream of the sensor or by using existing valves and
gates.
Normal operation valves 1 and 2 open
Zero point adjustment with pump pressure Valve 1 open / valve 2 closed
Zero point adjustment without pump pressure Valve 1 closed / valve 2 open
A0003601
Fig. 37: Zero point adjustment and shutoff valves
"
Caution!
The currently valid zero point value can be viewed using the "ZEROPOINT" function
(È "Description of Device Parameters" manual).
1
2

Table of Contents

Other manuals for Endress+Hauser Cubemass DCI

Related product manuals