Commissioning Cubemass DCI MODBUS RS485
54 Endress+Hauser
6.3.5 Data back-up/transfer
You can use the T-DAT SAVE/LOAD function to transfer data (device parameters and settings)
between the T-DAT (removable memory) and the EEPROM (device memory).
This is required for the following applications:
Creating a backup: current data are transmitted from an EEPROM to the T-DAT.
Replacing a transmitter: current data are copied from an EEPROM to the T-DAT, then transmitted
to the EEPROM of the new transmitter.
Duplicating data: current data are copied from an EEPROM to the T-DAT, then transmitted to
EEPROMs of identical measuring points.
!
Note!
Installing and removing the T-DAT ä 77.
A0001221-en
Fig. 31: Data storage/transmission with T-DAT SAVE/LOAD
Notes on the LOAD and SAVE options:
LOAD:
Data are transmitted from the T-DAT to the EEPROM.
!
Note!
Previously saved settings on the EEPROM are deleted.
This option is only available if the T-DAT contains valid data.
This selection can be made only if the software version of the T-DAT is the same or newer than
that of the EEPROM. Otherwise, the error message "TRANSM. SW-DAT" appears after the restart
and the LOAD function is subsequently no longer available.
SAVE:
Data are transmitted from the EEPROM to the T-DAT.
F
Esc
E
+
-
XXX.XXX.XX
F
F
F
F
F
F
P
P
P
P
N
O
T-DAT
SAVE/LOAD
Quick Setup
HOME
POSITION
LOAD
YES
NO
CANCELSAVE
YES
NO
Restart of the
measuring device
Input is
saved

Table of Contents

Other manuals for Endress+Hauser Cubemass DCI

Related product manuals