4.2 Connecting the remote version
4.2.1 Connecting the connecting cable for sensor/transmitter
#
Warning!
Risk of electric shock. Switch off the power supply before opening the device.
Do not install or wire the device while it is connected to the power supply.
Failure to comply with this precaution can result in irreparable damage to the electronics.
Risk of electric shock. Connect the protective ground to the ground terminal on the housing
before the power supply is applied.
You may only connect the sensor to the transmitter with the same serial number. Communication
errors can occur if this is not observed when connecting the devices.
1. Remove the cover (d) from the connection compartment or sensor housing.
2. Feed the connecting cable (e) through the appropriate cable runs.
3. Establish the wiring between the sensor and transmitter in accordance with the wiring diagram
( å 13 or wiring layout in the screw cap).
4. Seal the connection compartment or the transmitter housing again.
A0003681
Fig. 13: Connecting the remote version
a Transmitter wall-mount housing: non-hazardous area separate documentation
b Transmitter wall-mount housing: ATEX II2G / Zone 1 / NEC/CEC separate Ex documentation
cSensor connection housing
d Cover of connection compartment or connection housing
e Connecting cable
Terminal No.: 4/5 = gray; 6/7 = green; 8 = yellow; 9/10 = pink; 11/12 = white; 41/42 = brown
4.2.2 Cable specification, connecting cable
The specifications of the cable connecting the transmitter and the sensor of the remote version are
as follows:
6 × 0.38 mm
2
(20 AWG) PVC cable with common shield and individually shielded cores
Conductor resistance: 50 Ω/km ( 0.015 Ω/ft)
Capacitance core/shield: 140 pF/m ( 42.7 pF/ft)
Cable length: max. 20 m (65.6 ft)
Permanent operating temperature: max. +105 °C (+221 °F)
!
Note!
The cable must be installed securely, to prevent movement.
4 5 6 7 8 9 10 11 12 41 42
4 5 6 7 8 9 10 11 12 41 42
S1 S1 S2 S2 GND TM TM TT TT
++ ++
++ ++
S1 S1 S2 S2 GND TM TM TT TT
a
b
c
d
d
d
e