Wiring Cubemass DCI MODBUS RS485
24 Endress+Hauser
4.3.2 Terminal assignment
Electrical values for inputs ä 83.
Electrical values for outputs ä 84.
4.4 Degree of protection
The measuring device fulfills all the requirements for IP 67.
Compliance with the following points is mandatory following installation in the field or servicing,
in order to ensure that IP 67 protection is maintained:
The housing seals must be clean and undamaged when inserted into their grooves. The seals must
be dried, cleaned or replaced if necessary.
The screws and screw covers must be firmly tightened.
The cables used for connection must be of the specified outside diameter ä 85, Cable entries.
The cable entries must be firmly tightened (point a å 15).
The cable must loop down in front of the cable entry ("water trap")
(point b å 15). This arrangement prevents moisture penetrating the entry.
!
Note!
The cable entries may not point up.
A0001914
Fig. 15: Installation instructions, cable entries
Remove all unused cable entries and insert plugs instead.
Do not remove the grommet from the cable entry.
"
Caution!
Do not loosen the screws of the sensor housing, as otherwise the degree of protection guaranteed
by Endress+Hauser no longer applies.
Terminal No. (inputs/outputs)
Order version 20 (+) / 21 (-) 22 (+) / 23 (-) 24 (+) / 25 (-) 26 (+) / 27 (–)
Fixed communication boards (permanent assignment)
8CN**-**Q********* Status input MODBUS RS485
Flexible communication boards
8CN**-**N********* Current output Frequency output Status input MODBUS RS485
8CN**-**7********* Relay output 2 Relay output 1 Status input MODBUS RS485