EasyManua.ls Logo

Endress+Hauser Cubemass DCI - 10.1.6 Performance characteristics

Endress+Hauser Cubemass DCI
100 pages
To Next Page IconTo Next Page
To Next Page IconTo Next Page
To Previous Page IconTo Previous Page
To Previous Page IconTo Previous Page
Loading...
Technical data Cubemass DCI MODBUS RS485
86 Endress+Hauser
10.1.6 Performance characteristics
Reference operating
conditions
Error limits following ISO/DIS 11631
Water, typically 20 to 30 °C (68 to 86 °F); 2 to 4 bar (30 to 60 psi)
Data as per the calibration report ±5 °C (±9 °F) and ±2 bar (±30 psi)
Data on the measured error based on accredited calibration rigs traced back to ISO 17025
Maximum measured error The values indicated refer to the pulse/frequency output.
The additional measured error for the current output is typically ±5 μA.
Basis for calculations ä 88.
o.r. = of reading
Mass flow and volume flow (liquids)
±0.10% o.r. (mass flow)
±0.10% o.r. (volume flow)
Density (liquids)
±0.001 g/cc (after field density calibration or under reference conditions)
±0.002 g/cc (special density calibration (optional)
Calibration range: 0.0 to 2.0 g/cc, 5 to 80 °C (41 to 176 °F)
Application range: 0.0 to 5.0 g/cc, –50 to 200 °C (–58 to 392 °F)
±0.02 g/cc (standard calibration)
Temperature
±0.5 °C ± 0.005 · T °C
(±1.0 °F ± 0.003 · (T – 32) °F)
T = Fluid temperature
Zero point stability
DN Max. full scale value Zero point stability
[mm] [inch] [kg/h] [lb/min] [kg/h] [lb/min]
1 1/24" 0 to 20 0 to 0.75 0.0008 0.00003
2 1/12" 0 to 100 0 to 3.7 0.002 0.00007
4 1/8" 0 to 450 0 to 16.5 0.014 0.0005
6 ¼" 0 to 1000 0 to 37 0.02 0.0007

Table of Contents

Other manuals for Endress+Hauser Cubemass DCI

Related product manuals